CalDate
8/5/2024 12:00:00 AM
CalibratedBy
Ashtique
Recordings
Internal Jaw-40mm=40.00mm ; 80mm=80.00 mm, 120mm=120.00 mm; 150mm=150.00 mm External Jaw 20mm=20.00 mm;; 50mm=50.00 mm Depth 0mm=0mm ; 100mm=100.00 mm. Jaw alignment ok no damage, no corrosion, nicks, numbering is visible, screen output is good.
Notes
Fit for use